Clean Image Sheets
STEP 1, Daily Cleaning, Clean Image
Clean Crisp Film Every time.Eliminates Grainy Film.
• Process a new Clean Image Sheet as the first film of the day to remove artifacts that stick to rollers overnight.
• After each cleaning process a new Clean Image Sheet to remove any loosened debris.
• Replenish with Panopro Developer and Fixer.

Clear Image Radiographic Cleaner
STEP 2, Weekly Cleaning, Clear Image
Insures Quality X-Rays.
• Remove rollers and rinse well with warm water.
• Apply Clear Image to each roller or peri-type transport with the foaming applicator.
• Wait 2-3 minutes, scrub rollers with a green 3M Scotch-Brite pad and rinse well with warm water. For peri-type: use a stiff brush to clean the transport grooves.
• Return rollers or peri-type transport and replenish with Panopro Developer and Fixer.
• Process a new Clean Image
Sheet as the first film.
• Apply Clear Image monthly to fixer rollers, wash rack, and tanks.

Pro-32 Professional Strength Quarterly Cleaner
STEP 3, Monthly Cleaning, PRO-32
20-Min. Soak Evaporates Silver. No Odor, No Fumes, No Hazardous Shipping Fees.
• Rinse rollers and tanks with warm water.
• Prepare Pro-32 solution. Mix 3.5oz. per gallon of water in rubber tub or well-rinsed developer tank. Safety First: wear safety glasses, nitrile gloves, and long sleeves.
• Soak rollers or peri-type transports in Pro-32 solution for 20-45 minutes.
• Rinse and soak rollers or peri-type transports in fresh water for 10 minutes.
• Process a new Clean Image Sheet as the first film.

Panopro Developer and Fixer
Once Clean, Keep it clean
Once Clean, Keep Your Processor Running Cleaner Longer with Panopro Developer & Fixer.
Dark chemistry is a visible sign of silver buildup, oxidation, and poor quality x-rays.
Panopro stays fresh longer. Lighter chemistry is proof of a cleaner processor and sharper x-rays.